What is the most accurate wearable for SpO2?
Apple has the strongest current SpO2 evidence. Withings and Samsung share the second tier, while Garmin ranks fourth. Several other brands have useful validation data but not enough connected comparative evidence for a defensible numbered position.
SpO2 accuracy by brand
The ranking gives the most weight to same-protocol comparisons against clinical pulse oximetry or arterial blood, especially when oxygen saturation is deliberately lowered. Normal-range accuracy, low-oxygen accuracy and measurement availability are considered separately.
Rank | Brand | Verdict |
|---|---|---|
1 | Strongest overall evidence across controlled and pooled studies | |
2= | Very strong controlled-hypoxia performance, with availability caveats | |
2= | Very strong controlled evidence, including arterial-blood validation | |
4 | Mixed by model and condition; newer evidence is stronger | |
NR | Excellent altitude result, but comparative evidence is too thin | |
NR | Some clinical evidence, but too little connected comparison data | |
NR | Limited evidence and weak performance in one clinical population | |
NR | No independent peer-reviewed SpO2 validation strong enough to rank | |
NR | No independent peer-reviewed SpO2 validation strong enough to rank |
How the ranking was decided
SpO2 is highly condition-dependent. A watch can look excellent around normal saturation and become much less reliable when oxygen levels fall. For that reason, controlled hypoxia and arterial-blood comparisons carry more weight than ordinary resting measurements, and a device that fails to return a reading is not treated as accurate simply because the successful readings look good.
Apple takes first place because it is strongest across the connected evidence rather than because of one isolated study. In the three-watch controlled-hypoxia comparison, Apple Watch Series 8 had RMSE 2.2%, mean bias -0.1 percentage points and 93% hypoxemia-detection accuracy, ahead of both Withings and Samsung. The four-watch clinical comparison also put Apple closest to the reference, and the 2026 living meta-analysis pooled nine Apple Watch SpO2 studies with mean bias near zero.
Apple is not flawless at very low saturation. A 2026 arterial-blood controlled-hypoxia study found an overall Arms of 4.63%, rising to 5.82% below 88% saturation, with a tendency to overestimate low values. That is why the ranking describes Apple as the strongest consumer evidence rather than as a clinical replacement.
Withings and Samsung share second because the best direct comparison does not support a clean separation once the wider evidence is considered. In the same controlled-hypoxia study, Withings ScanWatch had RMSE 3.5% and 89% detection accuracy versus Samsung Galaxy Watch 5 at 3.8% and 84%, so Withings has the head-to-head edge. Samsung is then strengthened by a separate arterial-blood validation in which Galaxy Watch 4 achieved RMSD 1.8% across saturation from roughly 70% to 100%. A single-brand study cannot reverse the direct comparison, but it makes a shared tier more defensible than pretending the gap is settled.
The clinical evidence also shows why availability matters. In a prospective multi-watch study, Withings had substantial missing measurements even when its successful readings could be accurate. In a separate hospital COVID-19 study, Apple had higher overall classification accuracy while Withings had higher sensitivity for detecting low saturation. Those results support both brands while showing that the preferred device can depend on what failure mode matters most.
Garmin ranks fourth because it is the only other brand with enough connected evidence to place. The four-watch comparison found Garmin Venu 2S had much larger error than Apple, while Fenix 6 Pro failed to return a substantial share of measurements. Older high-altitude Fenix 5X Plus validation was also weak. A newer Fenix 7 activity study is more encouraging at rest and during recovery, but it was conducted in a separate round from Apple and cannot be used as a direct brand-to-brand numerical comparison.
Huawei stays unranked despite an impressive 2026 Watch GT 3 Pro altitude study, where RMSE remained below 1 percentage point across several elevations. It is a single-brand study with Huawei involvement and no direct bridge to the main comparison network, so under the locked methodology it strengthens confidence in that device without establishing a brand rank.
Fitbit and Xiaomi also remain unranked. Fitbit has useful clinical SpO2 evidence but little direct linkage to the ranked brands. The Mi Band 6 clinical study found poor SpO2 agreement in children and adolescents with severe cerebral palsy, which is important but too population-specific and too isolated to define the whole brand. Oura and WHOOP are unranked because the database does not yet contain independent peer-reviewed SpO2 validation that connects them to the comparative evidence.
Supporting studies
AppleApple Watch Series 6, Apple Watch Series 7, Apple Watch, Apple Watch Series 88 studies
Apple has the deepest connected SpO2 evidence. It leads the strongest same-protocol controlled-hypoxia comparison, performs well in another multi-watch clinical comparison, and is backed by a living meta-analysis. Accuracy still deteriorates in true hypoxemia, where low readings are harder for consumer wearables.
- Activity / structured October 30, 2026Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation
Apple Watch produced 99.6% of seated readings and 100% of cooldown readings, but only 0.4% during treadmill walking because its on-demand SpO2 measurement requires the wearer to remain still.
- Controlled hypoxia March 27, 2026Performance of Wearable Pulse Oximetry During Controlled Hypoxia Induction: Instrument Validation Study
Against arterial blood gas during controlled hypoxia, Apple Watch Series 7 had overall Arms 4.63% and Arms 5.82% when SaO2 was below 88%; it tended to overestimate low saturation.
- Mixed January 10, 2026The accuracy of Apple Watch measurements: a living systematic review and meta-analysis
Across nine studies, Apple Watch had pooled SpO2 mean bias -0.04 percentage points with limits of agreement -4.01 to 3.94; error and variability increased in hypoxic ranges.
- Clinical May 13, 2024Reliability of Continuous Vital Sign Monitoring in Post-Operative Patients Employing Consumer-Grade Fitness Trackers: A Randomised Pilot Trial
SpO2 performance was materially weaker than heart rate, with correlation 0.21-0.68 and RMSE 2.82-4.1 percentage points across devices.
- Clinical February 26, 2024Accuracy of Smartwatch Pulse Oximetry Measurements in Hospitalized Patients With Coronavirus Disease 2019
Apple Watch had 84.9% overall accuracy for detecting clinically significant hypoxia, with 34.8% sensitivity and 97.5% specificity.
- Rest / structured November 14, 2023Evaluation of Leading Smartwatches for the Detection of Hypoxemia: Comparison to Reference Oximeter
Apple Watch Series 8 led the three-watch controlled hypoxia comparison, with full-range RMSE 2.2%, mean bias -0.1 percentage points and 93% accuracy for detecting SpO2 below 90%.
- Mixed July 12, 2023Investigating the Accuracy of Blood Oxygen Saturation Measurements in Common Consumer Smartwatches
Apple Watch Series 7 was closest to the clinical reference, with MAE 2.2%, MDE -0.4% and RMSE 2.9%.
- Rest / structured 2022Blood Oxygen App on Apple Watch
Apple's validation study used Series 6 and reported agreement designed to meet consumer wellness accuracy requirements across induced desaturation.
WithingsWithings ScanWatch4 studies
Withings ScanWatch performs very well in the best three-watch controlled-hypoxia comparison and slightly edges Samsung there. Other clinical work shows a more mixed picture, including substantial missingness in one study, so it shares second rather than taking a clean standalone position.
- Clinical May 13, 2024Reliability of Continuous Vital Sign Monitoring in Post-Operative Patients Employing Consumer-Grade Fitness Trackers: A Randomised Pilot Trial
SpO2 performance was materially weaker than heart rate, with correlation 0.21-0.68 and RMSE 2.82-4.1 percentage points across devices.
- Clinical February 26, 2024Accuracy of Smartwatch Pulse Oximetry Measurements in Hospitalized Patients With Coronavirus Disease 2019
ScanWatch had 78.5% overall accuracy for detecting clinically significant hypoxia, with higher sensitivity than Apple at 68.5% but lower specificity at 80.8%.
- Rest / structured November 14, 2023Evaluation of Leading Smartwatches for the Detection of Hypoxemia: Comparison to Reference Oximeter
ScanWatch met the ISO threshold and slightly outperformed Galaxy Watch 5 in the same controlled hypoxia protocol, with full-range RMSE 3.5%, mean bias 0.4 percentage points and 89% hypoxemia-detection accuracy.
- Mixed July 12, 2023Investigating the Accuracy of Blood Oxygen Saturation Measurements in Common Consumer Smartwatches
Withings ScanWatch failed to return a reading in about 31% of attempts, the highest missingness of the four devices.
SamsungSamsung Galaxy Watch 6, Samsung Galaxy Watch 4, Samsung Galaxy Watch, Samsung Galaxy Watch 54 studies
Samsung shares second. Galaxy Watch 5 trails Withings slightly in the direct three-watch controlled-hypoxia study, but Galaxy Watch 4 has excellent separate arterial-blood validation. That extra evidence strengthens confidence without overriding the direct head-to-head result.
- Rest / structured May 22, 2026Reliability and Safety of Smartwatch Blood Pressure and Oxygen Saturation Measurements in Older Adults: Instrument Validation Study
Galaxy Watch 6 oxygen saturation showed moderate agreement with the reference (ICC 0.68).
- Controlled hypoxia February 20, 2025Evaluation of Pulse Oximetry Accuracy in a Commercial Smartphone and Smartwatch Device During Human Hypoxia Laboratory Testing
Galaxy Watch 4 achieved 1.8% RMSD versus arterial blood SaO2 across 247 paired measurements from 70-100% saturation, meeting the FDA/ISO threshold used in the study.
- Mixed September 1, 2024Performance of a Commercial Smart Watch Compared to Polysomnography Reference for Overnight Continuous Oximetry Measurement and Sleep Apnea Evaluation
Overnight smartwatch oximetry tracked clinically meaningful desaturation patterns but was not equivalent to polysomnography oximetry.
- Rest / structured November 14, 2023Evaluation of Leading Smartwatches for the Detection of Hypoxemia: Comparison to Reference Oximeter
Galaxy Watch 5 met the ISO accuracy threshold in controlled hypoxia, with full-range RMSE 3.8%, mean bias -2.6 percentage points and 84% accuracy for detecting SpO2 below 90%.
GarminGarmin Fenix 7, Garmin Venu, Garmin Fenix 6 Pro, Garmin Venu 2S, Garmin Fenix 5X Plus5 studies
Garmin has a broader but much more variable SpO2 record. Venu 2S performed poorly in a direct clinical comparison, Fenix 6 Pro had considerable missing data, and older altitude testing was weak. Newer Fenix 7 evidence is more encouraging at rest and during recovery.
- Activity / structured October 30, 2026Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation
Garmin produced all seated readings, 82.4% of treadmill readings and 98.4% of cooldown readings; it showed a small but significant -0.72 point signed bias during cooldown.
- Clinical July 16, 2026Accuracy of the Garmin Venu for measuring heart rate, respiratory rate, and peripheral oxygen saturation in patients hospitalized in a respiratory ward: a validation study
SpO2 was evaluated in clinical respiratory patients; this remains preliminary preprint evidence.
- Clinical May 13, 2024Reliability of Continuous Vital Sign Monitoring in Post-Operative Patients Employing Consumer-Grade Fitness Trackers: A Randomised Pilot Trial
SpO2 performance was materially weaker than heart rate, with correlation 0.21-0.68 and RMSE 2.82-4.1 percentage points across devices.
- Mixed July 12, 2023Investigating the Accuracy of Blood Oxygen Saturation Measurements in Common Consumer Smartwatches
Garmin Venu 2S showed the largest error among the tested devices, with MAE 5.8%, MDE 5.5% and RMSE 6.7%.
- Altitude September 23, 2021Validity of Peripheral Oxygen Saturation Measurements With the Garmin Fēnix 5X Plus Wearable Device at 4559 m
Fenix 5X Plus SpO2 showed poor-to-moderate agreement at altitude, with ICC about 0.55 and MAPE about 9.77%.
HuaweiHuawei Watch GT3 Pro1 study
Huawei Watch GT 3 Pro produced unusually strong results across altitude in a 2026 study. The study is single-brand and includes Huawei-affiliated authors, so it supports the device but does not provide the comparative bridge needed for a numbered brand rank.
- Altitude February 2, 2026Accuracy of wearable smartwatch for measuring blood pressure and oxygen saturation across a wide altitudinal gradient: a comparative study in Migrant and Resident populations
Watch GT3 Pro maintained low SpO2 RMSE across the altitude gradient, about 0.19% to 0.81%, within the study's ISO benchmark.
FitbitFitbit Sense 2, Fitbit Charge 42 studies
Fitbit has useful clinical SpO2 findings, including Charge 4 and Sense 2, but the current evidence is not connected to the main multi-brand controlled comparisons strongly enough to establish a relative rank.
- Clinical September 1, 2026Fitbit Sense 2 in Hospitalized General Medicine Patients: A Pilot Study Assessing Feasibility and Validity of Specific Measures
SpO2 was overestimated by about 1.7 percentage points and agreement was modest (ICC 0.47).
- Clinical November 1, 2024Validity of a Consumer-Based Wearable to Measure Clinical Parameters in Patients With Chronic Obstructive Pulmonary Disease and Healthy Controls: Observational Study
Oxygen-saturation agreement was poor in COPD and Fitbit tended to overestimate SpO2.
XiaomiXiaomi Mi Band 61 study
Xiaomi has very little SpO2 validation in the database. Mi Band 6 showed poor agreement in a clinical cerebral-palsy population, which is useful cautionary evidence but not enough to generalise across the brand.
- Clinical January 27, 2026Validity and Reliability of a Smart Band for Monitoring Cardiorespiratory Parameters in Children and Adolescents with Severe Cerebral Palsy
Mi Band 6 showed poor SpO2 agreement and high variability, with reliability ICC 0.55 and minimal detectable change 18.85%; the authors judged it unsuitable for accurate SpO2 assessment in this population.